Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.
Published December 17, 2014 · Recall initiated: September 23, 2014
Summary
FDA records medical device Class II recall Z-0577-2015, published December 17, 2014 and initiated September 23, 2014. Recalling firm: Invacare Corporation; Status: Terminated; Reported quantity: 5,963; Lot: The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Invacare Corporation
- Distributor
- Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
- Lot
- The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInvacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53March 13, 2024
- FDAAssist Bed Rail - 6632June 7, 2017
- FDAJunction Box - 1104343 and 1193560February 15, 2017
- FDABariatric Bed - BARPKG-IVC 1633February 15, 2017
- FDASyngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving andDecember 17, 2014
- FDAOxoid Legionella BCYE Growth Supplement packaged in packs of 10-100ml vials. A growth supplement for theDecember 17, 2014
Questions about this recall
Was Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. recalled?
Yes. FDA published recall Z-0577-2015 on December 17, 2014. Reported hazard: It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
What should I do if I have this product?
Voluntary: Firm initiated · Letter