QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
Published February 16, 2022 · Recall initiated: February 18, 2020
Summary
FDA records medical device Class II recall Z-0587-2022, published February 16, 2022 and initiated February 18, 2020. Recalling firm: Microbiologics Inc; Status: Terminated; Reported quantity: 10 units; Lot: Catalog Number: 5112P; Lot Numbers: 5112-01 (Expiration Date 07/31/2020), 5112-02 (Expiration Date 10/31/2020); UDI: 70845357026551. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Microbiologics Inc
- Distributor
- Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
- Lot
- Catalog Number: 5112P; Lot Numbers: 5112-01 (Expiration Date 07/31/2020), 5112-02 (Expiration Date 10/31/2020); UDI: 70845357026551
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAKWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from ATCC 17802, CatalogDecember 10, 2025
- FDALYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)August 6, 2025
- FDAKWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived fromJune 18, 2025
- FDAMicrobiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235;February 19, 2025
- FDABaxter MiniCap Extended Life PD Transfer Set with Twist Clamp, REF R5C4482EFebruary 16, 2022
- FDAGE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Architect system, nuclearFebruary 16, 2022
Questions about this recall
Was QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms recalled?
Yes. FDA published recall Z-0587-2022 on February 16, 2022. Reported hazard: Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
What should I do if I have this product?
Voluntary: Firm initiated · Letter