Stryker Howmedica Osteonics; Stryker T2 Humerus System Teflon Tube, sterile; REF 1806-0073S The T2 Nailing System is designed to treat complex fractures in Humeral, Femoral, Tibial and Ankle applications. The Teflon Tube is an optional instrument used to exchange the Ball Tip Guide Wire with the Smooth Tip Guide Wire once reaming is completed. Stryker Stryker Trauma GmbH 24232 Schonkirchen, Germany distributed in USA by: Howmedica Osteonics Corp., 235 Corporate Drive,
Published January 2, 2013 · Recall initiated: August 6, 2012
Summary
FDA records medical device Class II recall Z-0610-2013, published January 2, 2013 and initiated August 6, 2012. Recalling firm: Stryker Howmedica Osteonics Corp.; Status: Terminated; Reported quantity: 100 units (73 units US; 27 units OUS); Lot: Catalog No: 1806-0073S Lot Code:123596. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the manufacturing of the affected lot a manufacturing error has led to embrittlement of the tube material of the T2 Humerus Teflon Tube.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Howmedica Osteonics Corp.
- Distributor
- Worldwide distribution: USA (nationwide) including states of: AL, CO, CT, IL, IN, MO, MS, NC, NJ, NY, OR,and TX; and countries of: Australia, Canada, France, Spain and United Kingdom.
- Lot
- Catalog No: 1806-0073S Lot Code:123596
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDCMay 31, 2017
- FDASupracondylar Nail T2 SCN 09x300 mm T2 Supracondylar Nail System is intended to provide strong and stableSeptember 21, 2016
- FDAStryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used toAugust 17, 2016
- FDAStryker, T2 Guidewire, Ball-Tipped, 3 x 1000mm Product Usage: Intended for the temporary stabilization of boneMarch 2, 2016
- FDADiamigo i-Phone App The Diamigo app was intended for use as an educational tool in the management of diabetes.January 2, 2013
- FDABiosense Webster MobiCath Bi-Directional Guiding Sheath. Small Curve catalog no. D140010, Large Curve catalogJanuary 2, 2013
Questions about this recall
Was Stryker Howmedica Osteonics; Stryker T2 Humerus System Teflon Tube, sterile; REF 1806-0073S The T2 Nailing System is designed to treat complex fractures in Humeral, Femoral, Tibial and Ankle applications. The Teflon Tube is an optional instrument used to exchange the Ball Tip Guide Wire with the Smooth Tip Guide Wire once reaming is completed. Stryker Stryker Trauma GmbH 24232 Schonkirchen, Germany distributed in USA by: Howmedica Osteonics Corp., 235 Corporate Drive, recalled?
Yes. FDA published recall Z-0610-2013 on January 2, 2013. Reported hazard: During the manufacturing of the affected lot a manufacturing error has led to embrittlement of the tube material of the T2 Humerus Teflon Tube.
What should I do if I have this product?
Voluntary: Firm initiated · Letter