Puritan Bennett 840 ventilator systems software part number 4-070212-85, revision AB-AG. The intended use of the product is to provide mechanical ventilation to patients.

Published January 15, 2014 · Recall initiated: December 16, 2013

Summary

FDA records medical device Class I recall Z-0611-2014, published January 15, 2014 and initiated December 16, 2013. Recalling firm: Covidien; Status: Terminated; Reported quantity: 16,923 US, 39,372 OUS; Lot: Software Part number 4-070212-85, revision AB-AG. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Covidien is conducting a voluntary field corrective action on Puritan Bennett 840 ventilator systems with certain software revisions in response to customer reports of ventilator diagnostic code XB0069 in which the device stops mechanical ventilation due to a software error.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Covidien
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Austria, Estonia, Belgium, Spain, Italy, Poland, Bulgaria, Finland, Portugal, Switzerland, France, Lithuania, Romania, Cyprus, Great Britain, Czech Republic, Greece, Latvia, Slovenia, Germany, Hungary, Slovak Republic, Ireland, Netherlands, Turkey, Croatia, Andorra, Albania, Algeria, Bahrain, Bosnia and Herzegovina, Egypt, Georgia, Iran, Iraq, Israel, Jordan, Kenya, Kuwait, Lebanon, Libya, Macedonia, Montenegro, Morocco, Russia, Saudi Arabia, Serbia, South Africa, Syria, Sudan, Tunisia, UAE, Uganda, Puerto Rico, Mexico, Columbia, Chile, Brazil, Argentina, Uruguay, Panama, Costa Rica, Peru, China, Korea, Japan, Taiwan, Hong Kong, Malaysia, Singapore, Sri Lanka, Australia, New Zealand.
Lot
Software Part number 4-070212-85, revision AB-AG
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Puritan Bennett 840 ventilator systems software part number 4-070212-85, revision AB-AG. The intended use of the product is to provide mechanical ventilation to patients. recalled?

Yes. FDA published recall Z-0611-2014 on January 15, 2014. Reported hazard: Covidien is conducting a voluntary field corrective action on Puritan Bennett 840 ventilator systems with certain software revisions in response to customer reports of ventilator diagnostic code XB0069 in which the device stops mechanical ventilation due to a software error.

What should I do if I have this product?

Voluntary: Firm initiated · Letter