Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determine the objective refractive values and keratometric characteristics of the human eye. The results are used to assist the process of prescribing optical aids such as eyeglasses and contact lenses.
Published December 23, 2020 · Recall initiated: August 6, 2020
Summary
FDA records medical device Class II recall Z-0614-2021, published December 23, 2020 and initiated August 6, 2020. Recalling firm: Carl Zeiss Meditec AG; Status: Terminated; Reported quantity: 132 units; Lot: Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Carl Zeiss Meditec AG
- Distributor
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.
- Lot
- Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAZeiss Intrabeam Spherical Applicator 1.5 cm REF:304534-6000-557July 8, 2026
- FDAInSure ONE Test Strips (25 pack) vials-fecal immunochemical test (FIT) qualitatively detects human hemoglobinDecember 23, 2020
- FDARevMedX Trauma Dressings are packaged in a Tyvek Peel pouch, 1 Trauma Dressing per Tyvec Peel Pouch, 30December 23, 2020
- FDADePuy Synthes 12.0MM REAMER HEAD FOR RIA 2 STERILE -Orthopedic manual surgical instrument Part Code:December 23, 2020
- FDACarl Zeiss INTRABEAM PRS 500, product #304534-0000-500. The firm name on the label is Carl Zeiss Meditec AG,October 11, 2017
- FDATreatment packs used on the VisuMax Laser Keratome. Size M Manufactured by Carl Zeiss Meditec AG, Jena,March 19, 2014
Questions about this recall
Was Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determine the objective refractive values and keratometric characteristics of the human eye. The results are used to assist the process of prescribing optical aids such as eyeglasses and contact lenses. recalled?
Yes. FDA published recall Z-0614-2021 on December 23, 2020. Reported hazard: Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.
What should I do if I have this product?
Voluntary: Firm initiated · Letter