FDAFDA · Food and drugs·Recall number Z-0620-2025
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
Published December 11, 2024 · Recall initiated: October 31, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
- Lot
- Model Number: GIF-1TH190; UDI: 04953170343360, 04953170437014; Serial Number Range: Before 2446103 or After 2500453
- Risk classification
- Class II