Back to search
FDAFDA · Food and drugs·Recall number Z-0621-2013

Fisher & Paykel Healthcare reusable breathing circuit. Non-heated breathing circuit, 1.8m or 6 feet in length. Oxygen therapy delivery system for adult patient.

Published January 9, 2013 · Recall initiated: November 19, 2012

Summary

FDA records medical device Class I recall Z-0621-2013, published January 9, 2013 and initiated November 19, 2012. Recalling firm: Fisher & Paykel Healthcare, Ltd.; Status: Terminated; Reported quantity: 200 units total, 26 units US; Lot: Lots 110810 and 111020. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

FPH has become aware that one batch of tubes supplied to FPH may contain hole damage. These tubes were incorporated by FPH into certain 900MR068 reusable circuits with lot numbers 110810 and 111020. If this hole damage is not detected during the standard leak test before patient use, it could potentially result in a gas leak in the breathing system, which may lead to a loss of pressure for the int

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fisher & Paykel Healthcare, Ltd.
Distributor
Worldwide Distribution - USA including IL, IN, AR and GA.
Lot
Lots 110810 and 111020
Risk classification
Class I
Area (FDA)
device
Country
New Zealand

Related recalls

Questions about this recall

Was Fisher & Paykel Healthcare reusable breathing circuit. Non-heated breathing circuit, 1.8m or 6 feet in length. Oxygen therapy delivery system for adult patient. recalled?

Yes. FDA published recall Z-0621-2013 on January 9, 2013. Reported hazard: FPH has become aware that one batch of tubes supplied to FPH may contain hole damage. These tubes were incorporated by FPH into certain 900MR068 reusable circuits with lot numbers 110810 and 111020. If this hole damage is not detected during the standard leak test before patient use, it could potentially result in a gas leak in the breathing system, which may lead to a loss of pressure for the int

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail