BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263
Published December 11, 2024 · Recall initiated: November 6, 2024
Summary
FDA records medical device Class II recall Z-0621-2025, published December 11, 2024 and initiated November 6, 2024. Recalling firm: Becton Dickinson & Co.; Status: Ongoing; Reported quantity: 13,190 units; Lot: UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Co.
- Distributor
- Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.
- Lot
- UDi-DI: (01)30382902312636 Lot No. (Exp. Date): 2339360 (12/31/2024), 3010977 (01/31/25), 3058508 (03/31/2025), 3184064 (07/02/2025), 3234190 (08/21/2025).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.October 21, 2026
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.October 21, 2026
- FDABD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic useMay 6, 2026
- FDABD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic useMay 6, 2026
- FDAEVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190December 11, 2024
- FDAMEDLINE convenience kits labeled as: 1) TRACH TOTE, REF DYNJ85691; 2) TRACHEOSTOMY, REF DYNJ902123JDecember 11, 2024
Questions about this recall
Was BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263 recalled?
Yes. FDA published recall Z-0621-2025 on December 11, 2024. Reported hazard: BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail