The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnostic exams to be performed extraorally for improved clinical efficiency and patient experience. - Progeny Clarity Enhanced Filter Set provides superior diagnostic imaging quality - Exclusive Focal Trough Alignment reminder assists in consistent patient positioning - Progeny Imaging software included for download on an unlimited number of operatory workstations with no additional costs - Exclusive VantageTrust remote support and training service offered at no charge for the first 60 days of use - Ceph upgradeable for expanded extraoral exam capabilities - The Progeny Vantage Panoramic System s proudly designed and manufactured in the USA; 2013 Midmark Corporation; 675 Heathrow Drive Lincolnshire, IL 60069; 847-415-9800 Toll-free 888-924-3800
Published January 15, 2014 · Recall initiated: September 20, 2013
Summary
FDA records medical device Class II recall Z-0651-2014, published January 15, 2014 and initiated September 20, 2013. Recalling firm: Midmark Corp dba Progeny Inc; Status: Terminated; Reported quantity: 58 units; Lot: Vantage Panoramic X-Ray System firmware versions: 3.0, 3.1, 3.2, and 3.3. Vantage Panoramic X-Ray System Catalog Numbers and Serial Numbers: Catalog Number V5000, Serial Numbers ZM81025, ZM81026, ZM81030, ZZ08005, ZZ08006, ZZ08007, ZZ08008, ZZ08009, ZZ08010, ZZ08011, ZZ08012, ZZ08013, ZZ08015, ZZ08016, ZZ08017, Z08018, ZZ08019, ZZ08020, ZZ08021, ZZ08022, ZZ08023, ZB51001, ZB51002, ZB51006, ZB51007, ZT18018, ZY49014, ZK58019, ZS54011, ZP01043, ZV78016, ZE09006, ZN44018, ZS59007, ZS59001, ZW15007, ZW34003, ZW34012, ZW34018, ZY49012, ZY49015, ZZ02010, ZM81022, ZS54012, ZE09019, ZM81017, ZT18004, ZT18015, ZN44020, ZE09012, ZM81023, ZT27012; Catalog Number V5050, Serial Number ZZ08014; Catalog Number V5100, Serial Numbers ZM81027, ZM81028, ZZ08003; Catalog Number V5150, Serial Number ZZ08002; Catalog Number V5150C, Serial Number ZM81016. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. failure to stop").
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Midmark Corp dba Progeny Inc
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, CA, CO, CT, FL, GA, IL, KS, MA, NC, NH, NJ, OH, OK, PA, SC, TN, TX, VA, and WI, and the countries of Australia, Canada, and Russia.
- Lot
- Vantage Panoramic X-Ray System firmware versions: 3.0, 3.1, 3.2, and 3.3. Vantage Panoramic X-Ray System Catalog Numbers and Serial Numbers: Catalog Number V5000, Serial Numbers ZM81025, ZM81026, ZM81030, ZZ08005, ZZ08006, ZZ08007, ZZ08008, ZZ08009, ZZ08010, ZZ08011, ZZ08012, ZZ08013, ZZ08015, ZZ08016, ZZ08017, Z08018, ZZ08019, ZZ08020, ZZ08021, ZZ08022, ZZ08023, ZB51001, ZB51002, ZB51006, ZB51007, ZT18018, ZY49014, ZK58019, ZS54011, ZP01043, ZV78016, ZE09006, ZN44018, ZS59007, ZS59001, ZW15007, ZW34003, ZW34012, ZW34018, ZY49012, ZY49015, ZZ02010, ZM81022, ZS54012, ZE09019, ZM81017, ZT18004, ZT18015, ZN44020, ZE09012, ZM81023, ZT27012; Catalog Number V5050, Serial Number ZZ08014; Catalog Number V5100, Serial Numbers ZM81027, ZM81028, ZZ08003; Catalog Number V5150, Serial Number ZZ08002; Catalog Number V5150C, Serial Number ZM81016
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; NovaJanuary 15, 2014
- FDAArthrocare SPORTS MEDICINE, Knot Pusher, REF 25-3011, LOT #1049402, CON 1, 2013-10, NON-STERILE, Rx onlyJanuary 15, 2014
- FDACatalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) -January 15, 2014
- FDAProduct Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALertJanuary 15, 2014
- FDASOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE (Marketed As GE Healthcare CARESCAPE Patient DataJanuary 15, 2014
- FDA***REF BLACKMAX-N***Swivel/Angle Motor Assembly***Rx Only***The Anspach Effort, Inc. Intended for cutting andJanuary 15, 2014
Questions about this recall
Was The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnostic exams to be performed extraorally for improved clinical efficiency and patient experience. - Progeny Clarity Enhanced Filter Set provides superior diagnostic imaging quality - Exclusive Focal Trough Alignment reminder assists in consistent patient positioning - Progeny Imaging software included for download on an unlimited number of operatory workstations with no additional costs - Exclusive VantageTrust remote support and training service offered at no charge for the first 60 days of use - Ceph upgradeable for expanded extraoral exam capabilities - The Progeny Vantage Panoramic System s proudly designed and manufactured in the USA; 2013 Midmark Corporation; 675 Heathrow Drive Lincolnshire, IL 60069; 847-415-9800 Toll-free 888-924-3800 recalled?
Yes. FDA published recall Z-0651-2014 on January 15, 2014. Reported hazard: An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. failure to stop").
What should I do if I have this product?
Voluntary: Firm initiated · Letter