***REF BLACKMAX-N***Swivel/Angle Motor Assembly***Rx Only***The Anspach Effort, Inc. Intended for cutting and shaping bone.
Published January 15, 2014 · Recall initiated: April 25, 2013
Summary
FDA records medical device Class II recall Z-0654-2014, published January 15, 2014 and initiated April 25, 2013. Recalling firm: The Anspach Effort, Inc.; Status: Terminated; Reported quantity: 2,785; Lot: Serial Number 31496602. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Anspach Effort, Inc.
- Distributor
- Worldwide Distribution: US (nationwide) including Puerto Rico and states of: AL, AR, AZ, CA, CO, CT, DC, DE, GA, FL, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VE, VT, WA, WI, WY, and WV. Internationally to: AT, AU, BR, BS, CA, CH, CL, CO, CZ, CR, DE, DK, DO, EG, ES, FI, FR, GB, GR, HK, HU, ID, IL, IN, IS, JP, KH, KR, LB, LT, MX, MY, NL, NO, NZ, PA, PE, PH, PL, PT, SA, SG, SY, TH, TR, TW, and ZA.
- Lot
- Serial Number 31496602
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAElectric Handpiece EM Hand Control Product Usage: Electric systemMay 10, 2017
- FDA14cm QD Angle Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDAKeyless Driver Product Usage: Pneumatic systemMay 10, 2017
- FDA6.8cm Controlled Depth Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDAProduct Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALertJanuary 15, 2014
- FDA***REF XMAX***XMax Handpiece***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting andJanuary 15, 2014
Questions about this recall
Was ***REF BLACKMAX-N***Swivel/Angle Motor Assembly***Rx Only***The Anspach Effort, Inc. Intended for cutting and shaping bone. recalled?
Yes. FDA published recall Z-0654-2014 on January 15, 2014. Reported hazard: The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..
What should I do if I have this product?
Voluntary: Firm initiated · Letter