The TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM Ardis Inserter is intended for delivery of the TM Ardis Implant into the cleared disc space. The implant is secured to the inserter via finger-tightening the knob of the inserter shaft allowing the lateral grasping arms to secure the implant. Once the implant is inserted and its position confirmed radiographically, the inserter is removed from the implant by turning the shaft knob counterclockwise to loosen the lateral grasping arms from the implant.
Published January 16, 2013 · Recall initiated: September 20, 2012
Summary
FDA records medical device Class II recall Z-0659-2013, published January 16, 2013 and initiated September 20, 2012. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 170; Lot: Part number: 96-701-10001; Lot numbers: 61946711, 61946712, 61991588, 62000412,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer Spine is has received complaints of difficulty in turning the implant release knob of the TM Ardis Inserter after final positioning of the TM Ardis interbody spacer implant. There has been two reports where the inserter instrument was inadvertently removed from patient with the implant still engaged. The correction includes revised directions explaining the technique to remove the implant
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- USA nationwide; Europe, Sweden
- Lot
- Part number: 96-701-10001; Lot numbers: 61946711, 61946712, 61991588, 62000412,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAAutoPulse Resuscitation System Model 100, Mfg by:Zoll Circulation Sunnyvale, CA. The AutoPulse Model 100January 16, 2013
- FDAPT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder,January 16, 2013
Questions about this recall
Was The TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM Ardis Inserter is intended for delivery of the TM Ardis Implant into the cleared disc space. The implant is secured to the inserter via finger-tightening the knob of the inserter shaft allowing the lateral grasping arms to secure the implant. Once the implant is inserted and its position confirmed radiographically, the inserter is removed from the implant by turning the shaft knob counterclockwise to loosen the lateral grasping arms from the implant. recalled?
Yes. FDA published recall Z-0659-2013 on January 16, 2013. Reported hazard: Zimmer Spine is has received complaints of difficulty in turning the implant release knob of the TM Ardis Inserter after final positioning of the TM Ardis interbody spacer implant. There has been two reports where the inserter instrument was inadvertently removed from patient with the implant still engaged. The correction includes revised directions explaining the technique to remove the implant
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit