Back to search
FDAFDA · Food and drugs·Recall number Z-0660-2013

PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Published January 16, 2013 · Recall initiated: December 12, 2006

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Biomet, Inc.
Distributor
Worldwide Distribution -- US (nationwide) including the countries of , Canada, Japan, Argentina, Australia ,Netherlands, Europe and Chile.
Lot
Part 113950 Lot 093540, 093550,093570,093580,093590,093600,093610,093620,093720, 093730,093740,093760,093770,093780,282590,282600,282620,282630, 282680,282700,282710,489160,489180,489200,489210,489770,489780, 489790,714480,714550,714570,714700,714730,722210,722250,722280, 984770,984780,984790,984800,984810,984820,984830,984840, and 984850
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card