Back to search
FDAFDA · Food and drugs·Recall number Z-0662-2020

P28, Titan 3-D Wedge, Evans, Large, 10 mm, sterile, REF P03-EVN-1010-S, Type: Lagrge, Size: 10mm, (01) 00889795018182, Sterile R, Single Use Only

Published December 18, 2019 · Recall initiated: September 28, 2018

Summary

FDA records medical device Class II recall Z-0662-2020, published December 18, 2019 and initiated September 28, 2018. Recalling firm: Paragon 28, Inc.; Status: Terminated; Reported quantity: 64 Wedge Systems; Lot: Lot # BR4108. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It has been determined that the wedge implant kit is permanently mislabeled with the wrong size. The wedge implant is a 10mm size and is permanently marked as a 12mm size. Use of the product could result in improper preparation of the surgical site causing delays in surgery.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Paragon 28, Inc.
Distributor
AR, CA, CO, CT, DC, DE,FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MT, MS, NC, NE, NY, NJ, NV, OH, OR, PA, SC, SD, TN, TX, VT, WA, WI, WY OUS: Australia & Israel
Lot
Lot # BR4108
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was P28, Titan 3-D Wedge, Evans, Large, 10 mm, sterile, REF P03-EVN-1010-S, Type: Lagrge, Size: 10mm, (01) 00889795018182, Sterile R, Single Use Only recalled?

Yes. FDA published recall Z-0662-2020 on December 18, 2019. Reported hazard: It has been determined that the wedge implant kit is permanently mislabeled with the wrong size. The wedge implant is a 10mm size and is permanently marked as a 12mm size. Use of the product could result in improper preparation of the surgical site causing delays in surgery.

What should I do if I have this product?

Voluntary: Firm initiated · Letter