Back to search
FDAFDA · Food and drugs·Recall number Z-0675-2015

PKG, BLUNT FORCEPS, BLUNT, SERRATED, SPOON, P/N 0250080313. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Published December 24, 2014 · Recall initiated: November 17, 2014

Summary

FDA records medical device Class II recall Z-0675-2015, published December 24, 2014 and initiated November 17, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 180,573 units total; Lot: 43984 43987 44362 140498 140859 141064 246307 247408 340607 840366 840912 842117 842499 843735 844138 844601 845041 943743 944837 945749 1040109 1040579 1040937 1041623 1042256 1046590 1140173 1143127 1144355 1144488 1145337 1145813 11I0424 1240116 1240768 1240954 1241907 1242936 1244188 1244530 1246172 1340544 1343298 1343702 1344782 1344879 1344996 1345040 1345099 1346106 1346107 1346681 1440911 1443227. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Endoscopy
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lot
43984 43987 44362 140498 140859 141064 246307 247408 340607 840366 840912 842117 842499 843735 844138 844601 845041 943743 944837 945749 1040109 1040579 1040937 1041623 1042256 1046590 1140173 1143127 1144355 1144488 1145337 1145813 11I0424 1240116 1240768 1240954 1241907 1242936 1244188 1244530 1246172 1340544 1343298 1343702 1344782 1344879 1344996 1345040 1345099 1346106 1346107 1346681 1440911 1443227
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was PKG, BLUNT FORCEPS, BLUNT, SERRATED, SPOON, P/N 0250080313. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. recalled?

Yes. FDA published recall Z-0675-2015 on December 24, 2014. Reported hazard: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Voluntary: Firm initiated · Letter