Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)
Published December 18, 2024 · Recall initiated: October 25, 2024
Summary
FDA records medical device Class II recall Z-0679-2025, published December 18, 2024 and initiated October 25, 2024. Recalling firm: Noah Medical; Status: Ongoing; Reported quantity: 26; Lot: UDI-DI: 00850048825048. IFU Revision: J and all prior versions. All bronchoscope lots used with the following system serial numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB1540, CRB-1541, CRB-1543, CRB1554. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Noah Medical
- Distributor
- US Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
- Lot
- UDI-DI: 00850048825048. IFU Revision: J and all prior versions. All bronchoscope lots used with the following system serial numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB1540, CRB-1541, CRB-1543, CRB1554
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGalaxy System (GAL-001)December 25, 2024
- FDAGelato, Benzocaine 20% Topical Gel Anesthetic Gel, Net Wt. 1 oz. (30ml), Manufactured by Keystone IndustriesDecember 25, 2024
- FDAM&S Dental Supply Co LLC., Topical Anesthetic Gel, Benzocaine 20%, Net Wt. 1 oz. (30ml), Manufactured for: M&SDecember 25, 2024
- FDAHybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected areDecember 18, 2024
- FDADiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stoolDecember 18, 2024
- FDAGalaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories areMay 8, 2024
Questions about this recall
Was Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001) recalled?
Yes. FDA published recall Z-0679-2025 on December 18, 2024. Reported hazard: Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.
What should I do if I have this product?
Voluntary: Firm initiated · Letter