StatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use blood glucose ketone meter for near-patient testing

Published January 1, 2025 · Recall initiated: November 20, 2024

Summary

FDA records medical device Class I recall Z-0683-2025, published January 1, 2025 and initiated November 20, 2024. Recalling firm: Nova Biomedical Corporation; Status: Ongoing; Reported quantity: 1,615 units; Lot: Model No 63910. UDI-DI (01)1038548063910. Software Versions: v0.0.13.10 to v0.0.13.44. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Nova Biomedical Corporation
Distributor
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.
Lot
Model No 63910. UDI-DI (01)1038548063910. Software Versions: v0.0.13.10 to v0.0.13.44.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was StatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use blood glucose ketone meter for near-patient testing recalled?

Yes. FDA published recall Z-0683-2025 on January 1, 2025. Reported hazard: A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail