Galaxy System (GAL-001)
Published December 25, 2024 · Recall initiated: October 25, 2024
Summary
FDA records medical device Class II recall Z-0693-2025, published December 25, 2024 and initiated October 25, 2024. Recalling firm: Noah Medical; Status: Ongoing; Reported quantity: 26; Lot: UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB-1540, CRB-1541, CRB-1543, CRB-1554. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Noah Medical
- Distributor
- US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
- Lot
- UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB-1540, CRB-1541, CRB-1543, CRB-1554
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: K-Systems Product Name: Multiple/unknown Model/Catalog Number: Multiple/unknown Software Version:December 25, 2024
- FDAHUDSON RCI mBrace, ET Tube Holder and 2 Point Head Strap, REF DYNJMBRC2; tracheal tube fixation deviceDecember 25, 2024
- FDABrand Name: K-Systems Product Name: R65 Trolley Model/Catalog Number: K24012 Software Version: N/A ProductDecember 25, 2024
- FDAHUDSON RCI mBrace, ET Tube Holder with Bite Block and 2 Point Head Strap, REF DYNJMBRC2B; tracheal tubeDecember 25, 2024
- FDAGalaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the NoahDecember 18, 2024
- FDAGalaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories areMay 8, 2024
Questions about this recall
Was Galaxy System (GAL-001) recalled?
Yes. FDA published recall Z-0693-2025 on December 25, 2024. Reported hazard: Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
What should I do if I have this product?
Voluntary: Firm initiated · Letter