Galaxy System (GAL-001)
Published December 25, 2024 · Recall initiated: October 25, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Noah Medical
- Distributor
- US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
- Lot
- UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB-1540, CRB-1541, CRB-1543, CRB-1554
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States