Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.

Published January 30, 2013 · Recall initiated: November 19, 2012

Summary

FDA records medical device Class III recall Z-0696-2013, published January 30, 2013 and initiated November 19, 2012. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 845 Units; Lot: Part Numbers: 28065047001 and 03245233001. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus Analyzers. The Oracle TNS-Listener can be considered as an authentication and redirection component between the client and the database inside the Roche product. This means an attacker could register their own non-Roche database via remote access

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Roche Diagnostics Operations, Inc.
Distributor
Nationwide Distribution including AL, AR, AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR.
Lot
Part Numbers: 28065047001 and 03245233001
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Roche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer. recalled?

Yes. FDA published recall Z-0696-2013 on January 30, 2013. Reported hazard: A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus Analyzers. The Oracle TNS-Listener can be considered as an authentication and redirection component between the client and the database inside the Roche product. This means an attacker could register their own non-Roche database via remote access