ddRCombi Plus FP, Model Number 8.2000.0070.0 (US Version), 8.2000.0060.0 (Europe Version) The ddRCombi Plus FP is a fully digital, multifunctional X-ray system that allows the direct digitization of X-ray information without use of cassettes or other media. It can perform all applications in orthopedics and trauma as well as pulmonary imaging.

Published January 16, 2019 · Recall initiated: November 14, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A component of the device was recalled by a supplier. In rare cases of inadequate maintenance or very high clinical workload, the first steel wire rope of the lifting column, which is designed to take the load, can break without triggering the safety lock. This can lead to overloading of the safety cable, which is not designed to permanently carry the load of the lifting column or continuously move under load. This could cause material fatigue and cause the arm to drop unexpectedly during patient positioning. If the arm drops unexpectedly, patients and users may be seriously injured.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Swissray Medical AG
Distributor
Worldwide distribution: US distribution to states of: IL and MA and countries of: Czech Republic, Germany, Poland and Peru.
Lot
Serial Numbers: S403088, S403089, S403165
Risk classification
Class II
Area (FDA)
device
Country
Switzerland