Low Volume Cartridge IFU (SKU CAR-125-B) used with the following devices: 1) NxStage One, Model NX1000-1; 2) NxStage VersiHD Cycler, Model Numbers: NX1000-16, and NX1000-16-A; 3) NxStage One, Model Numbers: NX1000-3 and NX1000-3-A; 4) NxStage One, Model Numbers: NX1000-10 and NX1000-10-A;
Published December 25, 2019 · Recall initiated: August 27, 2019
Summary
FDA records medical device Class II recall Z-0697-2020, published December 25, 2019 and initiated August 27, 2019. Recalling firm: NxStage Medical, Inc.; Status: Terminated; Reported quantity: 1803 units; Lot: Product 1) UDI # - M535NX100010 Product 2) UDI # - M535NX1000160 and M535NX100016A0 Product 3) UDI # - M535NX100030 and M535NX10003A0 Product 4) UDI # - M535NX1000100 and M535NX100010A0 Lot Numbers: 60677021, 60877053, 70477033, 71177020, 80377002, 80477018, 80877011, 81077018, 81177047, 90277048 and 90677002. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- NxStage Medical, Inc.
- Distributor
- AL, AZ, CA, NC & VA
- Lot
- Product 1) UDI # - M535NX100010 Product 2) UDI # - M535NX1000160 and M535NX100016A0 Product 3) UDI # - M535NX100030 and M535NX10003A0 Product 4) UDI # - M535NX1000100 and M535NX100010A0 Lot Numbers: 60677021, 60877053, 70477033, 71177020, 80377002, 80477018, 80877011, 81077018, 81177047, 90277048 and 90677002
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAA1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062December 25, 2019
- FDANatrelle Saline-Filled Breast Implant, for the following Styles and sizes, Style LP, Style 68 HP, Style 68 MP,December 25, 2019
- FDANxStage PureFlow B Solution-RFP-402, Premixed Dialysate for Hemodialysis NxStage PureFlow B Solution isDecember 5, 2018
- FDANxStage PureFlow B Solution-RFP-400, Premixed Dialysate for Hemodialysis. NxStage PureFlow B Solution isDecember 5, 2018
- FDANxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5December 2, 2015
- FDANxStage Dialysate Sack (SAK) with Concentrate for Use with PureFlow SL REF: SAK-302 Lactate , 40, Potassium 1,July 2, 2014
Questions about this recall
Was Low Volume Cartridge IFU (SKU CAR-125-B) used with the following devices: 1) NxStage One, Model NX1000-1; 2) NxStage VersiHD Cycler, Model Numbers: NX1000-16, and NX1000-16-A; 3) NxStage One, Model Numbers: NX1000-3 and NX1000-3-A; 4) NxStage One, Model Numbers: NX1000-10 and NX1000-10-A; recalled?
Yes. FDA published recall Z-0697-2020 on December 25, 2019. Reported hazard: There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.
What should I do if I have this product?
Voluntary: Firm initiated · Letter