Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli.
Published January 30, 2013 · Recall initiated: December 13, 2012
Summary
FDA records medical device Class II recall Z-0699-2013, published January 30, 2013 and initiated December 13, 2012. Recalling firm: Becton Dickinson & Co.; Status: Terminated; Reported quantity: 67,820 units; Lot: Lot Number/ Exp Date 2086315 February 20, 2013 2101435 February 20, 2013 2104316 February 27, 2013 2129460 February 27, 2013 2129487 March 13, 2013 2163211 February 27, 2013 2163213 March 20, 2013 2163215 March 27, 2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Co.
- Distributor
- Worldwide Distribution-USA including the states of MA, TN, KY, IN, GA, NC, TX, NY, FL, AZ, WA, NJ, MI, and CA, and the countries of Uruguay, Taiwan, Singapore, Australia, South Korea, Mexico, Hong Kong, Guatemala, Colombia, Chile, Canada, India, Japan, Brazil, China and Belgium.
- Lot
- Lot Number/ Exp Date 2086315 February 20, 2013 2101435 February 20, 2013 2104316 February 27, 2013 2129460 February 27, 2013 2129487 March 13, 2013 2163211 February 27, 2013 2163213 March 20, 2013 2163215 March 27, 2013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.October 21, 2026
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.October 21, 2026
- FDABD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic useMay 6, 2026
- FDABD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic useMay 6, 2026
- FDAAbbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100January 30, 2013
- FDAGreenLight HPS (High Performance System) GreenLight XPS (Xcelerated Performance System) Manufactured by AMS,January 30, 2013
Questions about this recall
Was Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli. recalled?
Yes. FDA published recall Z-0699-2013 on January 30, 2013. Reported hazard: In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification.
What should I do if I have this product?
Voluntary: Firm initiated · Letter