Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use
Published February 6, 2013 · Recall initiated: December 13, 2012
Summary
FDA records medical device Class II recall Z-0702-2013, published February 6, 2013 and initiated December 13, 2012. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 16 units; Lot: Serial Numbers: 10110770 11030805 11070848 11070845 11070849 11030806 11090864 11110900 12020982 114087 12030999 12061020 12041003 12061014 12061018 12081040 :. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare Inc.
- Distributor
- US Nationwide Distribution including the states of AL, CA, CO, CT, IL, MD, MI, MO, NY, OH, PA
- Lot
- Serial Numbers: 10110770 11030805 11070848 11070845 11070849 11030806 11090864 11110900 12020982 114087 12030999 12061020 12041003 12061014 12061018 12081040 :
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMobile Diagnostic X-Ray System with Digital PanelApril 12, 2017
- FDAPhilips IntellVue Info Center iX 866023; PIIC iX Upgrade 866024; IntellVue Server iX Upgrade 866026; PIICJune 18, 2014
- FDAPhilips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.September 4, 2013
- FDAPhilips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage:August 28, 2013
- FDAMotorola Solutions PCS3000/CS3070 Laser Utility/Peripheral Laser ProductsFebruary 6, 2013
- FDABacT/ALERT SA Blood Culture Bottle. The BacT/ALERT SA reagent bottle is intended to be recover and detectFebruary 6, 2013
Questions about this recall
Was Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use recalled?
Yes. FDA published recall Z-0702-2013 on February 6, 2013. Reported hazard: Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21
What should I do if I have this product?
Voluntary: Firm initiated · Letter