FDAFDA · Food and drugs·Recall number Z-0702-2013

Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use

Published February 6, 2013 · Recall initiated: December 13, 2012

Summary

FDA records medical device Class II recall Z-0702-2013, published February 6, 2013 and initiated December 13, 2012. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 16 units; Lot: Serial Numbers: 10110770 11030805 11070848 11070845 11070849 11030806 11090864 11110900 12020982 114087 12030999 12061020 12041003 12061014 12061018 12081040 :. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Healthcare Inc.
Distributor
US Nationwide Distribution including the states of AL, CA, CO, CT, IL, MD, MI, MO, NY, OH, PA
Lot
Serial Numbers: 10110770 11030805 11070848 11070845 11070849 11030806 11090864 11110900 12020982 114087 12030999 12061020 12041003 12061014 12061018 12081040 :
Risk classification
Class II
Area (FDA)
device
Country
United States