pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).
Published January 22, 2014 · Recall initiated: June 10, 2013
Summary
FDA records medical device Class II recall Z-0713-2014, published January 22, 2014 and initiated June 10, 2013. Recalling firm: Radiometer America Inc; Status: Terminated; Reported quantity: 209 boxes/6 membranes; Lot: Part #: 942-042, Lot #: R0971. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Radiometer America Inc
- Distributor
- Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.
- Lot
- Part #: 942-042, Lot #: R0971
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAABL800 analyzer, model numbers 393-800 and 393-801.January 2, 2019
- FDAAutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+October 11, 2017
- FDAABL90 Flex Model #: 393-090 Portable, automated analyzer that measures pH, blood gases, electrolytes, glucose,March 22, 2017
- FDATCM CombiM Modules 903-111May 11, 2016
- FDABoston Scientific Extractor Pro XL Retrieval Balloon Catheter 12-15 mm Injects Below UPN :M00547140January 22, 2014
- FDADraeger Infinity Acute Care System Monitoring SolutionJanuary 22, 2014
Questions about this recall
Was pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen). recalled?
Yes. FDA published recall Z-0713-2014 on January 22, 2014. Reported hazard: RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.
What should I do if I have this product?
Voluntary: Firm initiated · Letter