PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870
Published January 24, 2024 · Recall initiated: August 29, 2023
Summary
FDA records medical device Class II recall Z-0726-2024, published January 24, 2024 and initiated August 29, 2023. Recalling firm: Limacorporate S.p.A; Status: Ongoing; Reported quantity: 179; Lot: Product Code/UDI-DI: 1975.14.500/08033390244824, 1975.14.800/08033390244893, 1975.14.510/08033390231299, 1975.14.810/08033390244909, 1975.14.615/08033390244848, 1975.14.815/08033390244916, 1975.14.865/08033390244930, 1975.14.620/08033390231312, 1975.14.820/08033390244923, 1975.14.870/08033390244947. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Limacorporate S.p.A
- Distributor
- US: IL, MO, CA, IN, MI, OK
- Lot
- Product Code/UDI-DI: 1975.14.500/08033390244824, 1975.14.800/08033390244893, 1975.14.510/08033390231299, 1975.14.810/08033390244909, 1975.14.615/08033390244848, 1975.14.815/08033390244916, 1975.14.865/08033390244930, 1975.14.620/08033390231312, 1975.14.820/08033390244923, 1975.14.870/08033390244947
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy
Related recalls
- FDASpectral CT 7500, Software Version 5.0.0.X. Computed Tomography X-Ray System.January 24, 2024
- FDAUltrasound EA (Ergonomic Advantage) table versions; G2 (GENERAL 2) -An Ultrasound examination tableJanuary 24, 2024
- FDAVasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3500.January 24, 2024
- FDADelta Cup Beater Positioner - Aligner, model # 9057.20.555 Used during implantation of the Delta AcetabularNovember 15, 2017
- FDASMR glenosphere impactor/extractor; Used to implant the glenosphere in the SMR Reverse Shoulder replacement.June 1, 2016
- FDASMR Shoulder/Spalla Trial Heads ECC .4 mm Adaptor, REF 9013.30.021; SMR Shoulder/Spalla Trial Heads ECC .8 mmSeptember 9, 2015
Questions about this recall
Was PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870 recalled?
Yes. FDA published recall Z-0726-2024 on January 24, 2024. Reported hazard: Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit