Lotus Edge Valve System, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System, 23mm, sterile, REF H749LVSUS230, GTIN 08714729960904
Published January 20, 2021 · Recall initiated: November 17, 2020
Summary
FDA records medical device Class I recall Z-0728-2021, published January 20, 2021 and initiated November 17, 2020. Recalling firm: Boston Scientific Corporation; Status: Terminated; Reported quantity: 219 devices; Lot: GTIN 08714729960904. All unexpired lot numbers, expiration dates 2-Nov-2020 through 29-Oct-2021. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- Distribution was made to AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. Government distribution was also made. Foreign distribution was made to Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Poland, Portugal, South Korea, Spain, Sweden, and Switzerland.
- UPC
- 08714729960904
- Lot
- GTIN 08714729960904. All unexpired lot numbers, expiration dates 2-Nov-2020 through 29-Oct-2021
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;October 21, 2026
- FDAImager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5;September 30, 2026
- FDABoston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVISeptember 23, 2026
- FDALUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model NumberSeptember 9, 2026
- FDAArtis zee/ zeego & Artis Q/ Q.zen Artis PhenoJanuary 20, 2021
- FDASOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge PlusJanuary 20, 2021
Questions about this recall
Was Lotus Edge Valve System, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System, 23mm, sterile, REF H749LVSUS230, GTIN 08714729960904 recalled?
Yes. FDA published recall Z-0728-2021 on January 20, 2021. Reported hazard: Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 08714729960904.