20/30 Priority Pack Accessory Kit/.096 RHV, REF: 1000186, or use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation catheters, atherectomy devices, stent delivery systems, intravascular ultrasound devices).
Published January 1, 2025 · Recall initiated: November 6, 2024
Summary
FDA records medical device Class II recall Z-0731-2025, published January 1, 2025 and initiated November 6, 2024. Recalling firm: Abbott Vascular Inc; Status: Ongoing; Reported quantity: 3,673; Lot: UDI-DI: 08717648013614, Lots: 60538750, 60538751, 60539113, 60545329. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Vascular Inc
- Distributor
- US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.
- Lot
- UDI-DI: 08717648013614, Lots: 60538750, 60538751, 60539113, 60545329
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA20/30 Priority Pack Accessory Kit w/Copilot, REF: 1003327, used for cardiovascular procedures.January 1, 2025
- FDAKIT,SPINE LH Convenience kit, SKU DYNJ906448A.January 1, 2025
- FDAPanther Fusion SARS-CoV-2/Flu A/B/RSV Assay, REF: PRD-07400January 1, 2025
- FDAepoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for useJanuary 1, 2025
- FDAStatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use bloodJanuary 1, 2025
- FDAepoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for useJanuary 1, 2025
Questions about this recall
Was 20/30 Priority Pack Accessory Kit/.096 RHV, REF: 1000186, or use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation catheters, atherectomy devices, stent delivery systems, intravascular ultrasound devices). recalled?
Yes. FDA published recall Z-0731-2025 on January 1, 2025. Reported hazard: Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
What should I do if I have this product?
Voluntary: Firm initiated · Letter