PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547
Published January 24, 2024 · Recall initiated: November 16, 2023
Summary
FDA records medical device Class II recall Z-0736-2024, published January 24, 2024 and initiated November 16, 2023. Recalling firm: Becton Dickinson & Company; Status: Ongoing; Reported quantity: 5,085,120 units; Lot: UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Company
- Distributor
- US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.
- Lot
- UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD®Spinal Tray with BD®Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDATECNIS Toric II OptiBlue IOL Models ZCWJanuary 24, 2024
- FDAV. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18January 24, 2024
Questions about this recall
Was PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547 recalled?
Yes. FDA published recall Z-0736-2024 on January 24, 2024. Reported hazard: An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
What should I do if I have this product?
Voluntary: Firm initiated · Letter