ABX PENTRA Uric Acid CP ref. A11A01670 is an in vitro diagnostic assay for the quantitative determination of uric acid in human serum, plasma and urine based on the enzymatic determination of uric acid using a chromogenic system in the presence of peroxidase and uricase (Trinder method).
Published February 10, 2016 · Recall initiated: January 11, 2016
Summary
FDA records medical device Class II recall Z-0743-2016, published February 10, 2016 and initiated January 11, 2016. Recalling firm: Horiba Instruments Inc; Status: Terminated; Reported quantity: 216; Lot: All Lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Horiba Instruments Inc
- Distributor
- U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
- Lot
- All Lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASyrSpend SF, a convenient, sweetened, sugar-free syrup-suspending vehicle specially formulated to assist inFebruary 17, 2016
- FDAABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnosticFebruary 10, 2016
- FDAABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determinationFebruary 10, 2016
- FDABacT/ALERT® SN, bioM®rieux, Inc. Product Usage: BacT/ALERT® Culture Bottles are used with the BacT/ALERT®February 10, 2016
- FDAHemosIL Plasminogen, Part No. 0020009000, IVD. Instrumentation Laboratory Company. Kit contents: 2 x 2.5 mLFebruary 10, 2016
- FDALite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D TheDecember 31, 2014
Questions about this recall
Was ABX PENTRA Uric Acid CP ref. A11A01670 is an in vitro diagnostic assay for the quantitative determination of uric acid in human serum, plasma and urine based on the enzymatic determination of uric acid using a chromogenic system in the presence of peroxidase and uricase (Trinder method). recalled?
Yes. FDA published recall Z-0743-2016 on February 10, 2016. Reported hazard: N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
What should I do if I have this product?
Voluntary: Firm initiated · Letter