Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45, 1011649-55, 1011649-65, 1011649-75, 1011650-55, 1011650-65, 1011651-45, 1011651-55. The RX Accunet EPS is indicated for use as a guide wire and embolic protection system.
Published February 6, 2013 · Recall initiated: November 30, 2011
Summary
FDA records medical device Class II recall Z-0757-2013, published February 6, 2013 and initiated November 30, 2011. Recalling firm: Abbott Vascular; Status: Terminated; Reported quantity: 1253 units (962 units in the US & Puerto Rico); Lot: Lot Number: 1081061, 1082561, 1091361, 1080561, 1081761, 1090261, 1091561, 1100461, 1080361, 1081661, 1090161, 1090861, 1092361, 1080861, 1082661, 1100761, 1081761, 1092861, 1082461, 1081161, 1082261, 1083061, 1081961, 1093061. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The recall was initiated because Abbot Vascular has discovered that the identified lots of the RX Accunet Embolic Protection System may exhibit difficult removal of the peel away sheath due to higher than normal wall thickness.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Vascular
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Austria, Belgium, China, Germany, Spain, France, United Kingdom, Italy, Pakistan, Poland, Russia, Slovakia, Taiwan, and South Africa.
- Lot
- Lot Number: 1081061, 1082561, 1091361, 1080561, 1081761, 1090261, 1091561, 1100461, 1080361, 1081661, 1090161, 1090861, 1092361, 1080861, 1082661, 1100761, 1081761, 1092861, 1082461, 1081161, 1082261, 1083061, 1081961, 1093061.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPlus 30 Priority Pack w/.096 RHV - REF 1000183 WPL2122277-01 (20-03-01) Plus 30 Priority Pack w/.115 RHV - REFApril 20, 2022
- FDANC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RXFebruary 26, 2020
- FDANC TRAVELER Coronary Dilatation Catheter, REF 1013157-15, Rx Only,NC Traveler RX 4.0 X 15MM The NC Traveler RXFebruary 26, 2020
- FDANC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RXFebruary 26, 2020
- FDABD Vacutainer Buffered Sodium Citrate (9NC) Plus Blood Collection Tubes; Sterile; IVD; REF 363083; 13 X 75 mm,February 6, 2013
- FDAEthicon Endo-Surgery Inc. The ENSEAL® G2 Tissue Sealers are indicated for bipolar coagulation and mechanicalFebruary 6, 2013
Questions about this recall
Was Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45, 1011649-55, 1011649-65, 1011649-75, 1011650-55, 1011650-65, 1011651-45, 1011651-55. The RX Accunet EPS is indicated for use as a guide wire and embolic protection system. recalled?
Yes. FDA published recall Z-0757-2013 on February 6, 2013. Reported hazard: The recall was initiated because Abbot Vascular has discovered that the identified lots of the RX Accunet Embolic Protection System may exhibit difficult removal of the peel away sheath due to higher than normal wall thickness.
What should I do if I have this product?
Voluntary: Firm initiated · Letter