COBAS TaqMan Mycobacteria tuberculosis (MTB) Test FOR IN VITRO DIAGNOSTIC USE.
Published February 6, 2013 · Recall initiated: November 28, 2012
Summary
FDA records medical device Class II recall Z-0762-2013, published February 6, 2013 and initiated November 28, 2012. Recalling firm: Roche Molecular Systems, Inc.; Status: Terminated; Reported quantity: 335 CE-IVD TDFs; 48 JPN-IVD TDFs; Lot: Material Number: CE-IVD: 04803531190 Kit, 05574455190 TDF; JPN-IVD: 04653360190 Kit, 05574447190 TDF. Lot/Serial Number: CE-IVD: All Kits, JPN-IVD: All Kits. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The COBAS TaqMan MTB Test Definition Files (TDF) v3.0, for both CE-IVD and JPN-IVD tests, being used with AMPLILINK software version 3.3 series and COBAS TaqMan MTB Test Kits, both CE-IVD and JPN-IVD, have an incorrect Target Elbow Max cutoff value of 48 creating the potential for false negative results to be generated for patient samples with Target Elbows in the range of 48 to 50.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Roche Molecular Systems, Inc.
- Distributor
- Worldwide Distribution - Foreign (22 EU countries - Austria, Belgium, Bulgaria, Croatia, Cypress, Czech Republic, Denmark, Germany, Finland, France, Greenland, Hungary, Iceland, Italy, Luxembourg, Netherland, Norway, Sweden, Switzerland, Spain, Turkey, UK); (41 Rest of World countries - See Attachment for full country distribution + Japan, Portugal and Mexico)
- Lot
- Material Number: CE-IVD: 04803531190 Kit, 05574455190 TDF; JPN-IVD: 04653360190 Kit, 05574447190 TDF. Lot/Serial Number: CE-IVD: All Kits, JPN-IVD: All Kits.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAcobas Liat Analyzer: respiratory virus panel nucleic acid assay system Catalog Number: 07341920190April 21, 2021
- FDANucleic Acid test intended for screening for MRSA/SA. Each kit has 20 tubes. in vitro diagnostic.April 4, 2018
- FDAcobas® Liat® AnalyzerJuly 27, 2016
- FDAcobas® PCR Urine Kit 100 PKT IVD cobas® PCR Female Swab Kit 100 PKT IVD cobas® PCR media 100T IVD cobas® PCRMarch 5, 2014
- FDARectal Retractors, titanium, item # AL07368000, a component of Varian Medical Ring and Tandem Applicator Sets;February 6, 2013
- FDAInfinity Enteralite Ambulatory Feeding Pumps, List number: INFKIT2 - Addendum to Operator's Manual ProductFebruary 6, 2013
Questions about this recall
Was COBAS TaqMan Mycobacteria tuberculosis (MTB) Test FOR IN VITRO DIAGNOSTIC USE. recalled?
Yes. FDA published recall Z-0762-2013 on February 6, 2013. Reported hazard: The COBAS TaqMan MTB Test Definition Files (TDF) v3.0, for both CE-IVD and JPN-IVD tests, being used with AMPLILINK software version 3.3 series and COBAS TaqMan MTB Test Kits, both CE-IVD and JPN-IVD, have an incorrect Target Elbow Max cutoff value of 48 creating the potential for false negative results to be generated for patient samples with Target Elbows in the range of 48 to 50.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail