Back to search
FDAFDA · Food and drugs·Recall number Z-0791-2014

Brilliance iCT, Brilliance 64 and Brilliance Big Bore The Brilliance iCT, Brilliance 64 and Brilliance Big Bore are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Published February 5, 2014 · Recall initiated: March 7, 2011

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The DoseRight feature suggest a mAs based on the measured patient size, a reference size and a reference mAs. When scanning large children, the suggested mAs may be higher than clinicians would expect.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution - US Distribution including the states of DC, IL, IN, MI, MO, NY, OH, OR, PA, TX and VT., and the countries of Australia, Belgium, China, Germany, India, Israel, Italy and The Netherlands.
Lot
Brilliance iCT: Model Number 728311 and 728306, Serial Number: 100103, 100019, 100023, 100170, 200047, 200013 200005 and 100087. Brilliance 64: Model Number 728231 and 728326, Seial Number: 4003, 9745, 9875, 90135, 90169, 95414, 300003, 300004, 300005 and 300010. Brilliance CT Big Bore, Model Number: 728243, Serial Number: 7006, 7060, 7154, 7298 and 7349.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card