GE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typical users of this system are authorized healthcare professionals. Centricity Universal Viewer is intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information.
Published February 13, 2019 · Recall initiated: December 20, 2018
Summary
FDA records medical device Class II recall Z-0793-2019, published February 13, 2019 and initiated December 20, 2018. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 297 (131 US; 166 OUS); Lot: Centricity Universal Viewer versions 6.0 SP4 or higher with Breast Imaging integrated with either IDI Mammo Workstation version 4.7.x or any version of SenoIris. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When switching back & forth between multiple UV instances in the Windows taskbar, the patient images displayed on the mammo high resolution monitors may not show images of the patient selected on UV, which could result in an incorrect diagnosis.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution: US (nationwide) in states of: AZ, CA, CO, DE, FL, IA, IL, IN, KS, MA, ME, MI, MN, MO, MS, MT, NE, NJ, NV, NY, OH, PA, PR, SC, SD, TN, TX, VA, VT, and WI; and countries of: Antigua, Australia, Austria, Belgium, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Cote d'Ivoire, France, Germany, Guatemala, India, Italy, Korea, Kuwait, Lebanon, Malaysia, Malta, Mexico, Panama, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot
- Centricity Universal Viewer versions 6.0 SP4 or higher with Breast Imaging integrated with either IDI Mammo Workstation version 4.7.x or any version of SenoIris.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAPortex 5ml Normal Saline (0.9%) Unit Dose®¿¿¿ ¿February 13, 2019
- FDAAscension Modular Radial Head (MRH) Radial Head Implant, Catalog Number, MRH-350-04-WWFebruary 13, 2019
Questions about this recall
Was GE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that displays medical images and data from various imaging sources, and from other healthcare information sources. Medical images and data can be displayed, communicated, stored, and processed. Typical users of this system are authorized healthcare professionals. Centricity Universal Viewer is intended to assist in the viewing, analysis, diagnostic interpretation, and sharing of images and other information. recalled?
Yes. FDA published recall Z-0793-2019 on February 13, 2019. Reported hazard: When switching back & forth between multiple UV instances in the Windows taskbar, the patient images displayed on the mammo high resolution monitors may not show images of the patient selected on UV, which could result in an incorrect diagnosis.
What should I do if I have this product?
Voluntary: Firm initiated · Letter