The Lumenis VersaCut Tissue Morcellator Product Usage: is intended for use under direct or endoscopic visualization for the morcellation and removal of dissected tissue during pelviscopic, laparoscopic, percutaneous, and open surgical procedures whenever access to the surgical site is limited.
Published February 20, 2013 · Recall initiated: January 2, 2013
Summary
FDA records medical device Class I recall Z-0795-2013, published February 20, 2013 and initiated January 2, 2013. Recalling firm: Lumenis Limited; Status: Terminated; Reported quantity: 931 units; Lot: Model Numbers 0637-245-01 (starter kit) and 0636-470-01 (Control Box) which includes all products since initial product release 1998. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Lumenis Ltd is recalling certain models of VersaCut Morcellator devices to perform corrective labeling as a result of three reported events of air embolism secondary to incorrect reversal tubing system hook-up.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Lumenis Limited
- Distributor
- Worldwide Distribution - US Nationwide including: CA, WI, NC, FL, PA, IN, NM, NJ, IA, WA, OH, FL, MD, MA, WY, NY, AZ, NH, MN, TN, OK, KY, TX, AK, MI, VT. Internationally: Japan, Nepal, India, UK, South Korea, Netherlands, Germany, Australia, China, Canada, Turkey, Spain, Italy, Egypt, Mexico, Bulgaria, Utd. Arab, Emir., Poland, Kuwait, Hong Kong, Saudi Arabia, Yemen, Qatar, Czech Republic, New Zealand, Malaysia, Portugal, France, Argentina, Singapore, Romania, Chile, Colombia, Switzerland, Bahrain, Brazil, Serbia/Montenero, Taiwan, Israel, Indonesia, Peru, Pakistan, Nepal, Belarus, Latvia, Sri Lanka, Russian Fed, Thailand, Azerbaijan, Belgium, and Denmark.
- Lot
- Model Numbers 0637-245-01 (starter kit) and 0636-470-01 (Control Box) which includes all products since initial product release 1998.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- Israel
Related recalls
- FDASt. Jude Medical, AGA Medical corporation, AMPLATZER® TorqVue® FX Delivery System. Model #'s: 9-ITVFX06F45/60,February 20, 2013
- FDAMindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, andFebruary 20, 2013
- FDAGiraffe and Panda Bag and Mask Resuscitation System The T-piece Resuscitation System [ventilator, emergencyFebruary 20, 2013
- FDAFresenius Liberty Cycler Set Single Conn. Ext. DL. Part Number: 050-87216. Acute and chronic peritonealFebruary 20, 2013
- FDAGE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator,February 20, 2013
- FDABacterin International, Inc., Osteoselect Demineralized Bone Matrix DMB Putty; Graft ID# B090056-664 Donor#February 20, 2013
Questions about this recall
Was The Lumenis VersaCut Tissue Morcellator Product Usage: is intended for use under direct or endoscopic visualization for the morcellation and removal of dissected tissue during pelviscopic, laparoscopic, percutaneous, and open surgical procedures whenever access to the surgical site is limited. recalled?
Yes. FDA published recall Z-0795-2013 on February 20, 2013. Reported hazard: Lumenis Ltd is recalling certain models of VersaCut Morcellator devices to perform corrective labeling as a result of three reported events of air embolism secondary to incorrect reversal tubing system hook-up.
What should I do if I have this product?
Voluntary: Firm initiated · Letter