FDAFDA · Food and drugs·Recall number Z-0800-2014
Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.
Published January 29, 2014 · Recall initiated: December 17, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution. USA nationwide and the country of Canada.
- Lot
- ADVIA 1200 with software versions: V2.00 and V2.01 ADVIA 1650 with software versions V3.52 and V4.01
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States