Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.

Published January 29, 2014 · Recall initiated: December 17, 2013

Summary

FDA records medical device Class II recall Z-0800-2014, published January 29, 2014 and initiated December 17, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 1563; Lot: ADVIA 1200 with software versions: V2.00 and V2.01 ADVIA 1650 with software versions V3.52 and V4.01. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide Distribution. USA nationwide and the country of Canada.
Lot
ADVIA 1200 with software versions: V2.00 and V2.01 ADVIA 1650 with software versions V3.52 and V4.01
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine. recalled?

Yes. FDA published recall Z-0800-2014 on January 29, 2014. Reported hazard: An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.

What should I do if I have this product?

Voluntary: Firm initiated · Letter