BriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, Qty 10, (01) 15055788722783 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.

Published January 22, 2020 · Recall initiated: November 27, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Flexicare Medical Ltd.
Distributor
US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
Lot
Part Number/Lot Number 040-722U 181000115 I CCW 040-722U 181001520 I CCW 040-722U 181200006 I CCW 040-722U 181202055 I CCW 040-722U 190100006 I CCW
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom