BriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF: 040-342U, Sterile EO, Qty 10, UDI: (01)15055788724053(17)230901(10)181002320
Published January 22, 2020 · Recall initiated: November 27, 2019
Summary
FDA records medical device Class II recall Z-0804-2020, published January 22, 2020 and initiated November 27, 2019. Recalling firm: Flexicare Medical Ltd.; Status: Terminated; Reported quantity: 1,024,890 devices; Lot: Part Number/Lot Number: 040-342U 180802596 I OGH 040-342U 181002320 I OGH 040-342U 181100085 I OGH 040-342U 181200025 I OGH. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Flexicare Medical Ltd.
- Distributor
- US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
- Lot
- Part Number/Lot Number: 040-342U 180802596 I OGH 040-342U 181002320 I OGH 040-342U 181100085 I OGH 040-342U 181200025 I OGH
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDABriteBlade Pro Single-Use Fiber Optic Handle and Blade Mac 3, CE 0120, Do not Re-use, Rx Only, REF, 040-333U,January 22, 2020
- FDABriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, QtyJanuary 22, 2020
- FDABriteBlade Pro Single-Use Fiber Optic Mac 3, CE01120, Do Not Re-use, Rx Only, REF 040-713U, Sterile EO, QtyJanuary 22, 2020
- FDASiemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: AngiographyJanuary 22, 2020
- FDAMicrobiologics QC Sets and Panels, Gram-Positive Blood Culture Control Panel (Inactivated Pellet), Catalog #January 22, 2020
- FDADawson-Mueller Drainage Catheter, Multipurpose Drainage Catheter, RPN UL T6.3-35-15-P-5S-CLDM-HC, UL T6.January 22, 2020
Questions about this recall
Was BriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF: 040-342U, Sterile EO, Qty 10, UDI: (01)15055788724053(17)230901(10)181002320 recalled?
Yes. FDA published recall Z-0804-2020 on January 22, 2020. Reported hazard: The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit