Aesculap(R) - Miethke Shunt System, FT040T - Control Reservoir Set w/Distal Catheter and FT075P 0 Ventric Cathet, w/Stylet and RT-ANG Guide Central Nervous System Shunt and Components intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
Published February 20, 2013 · Recall initiated: June 4, 2012
Summary
FDA records medical device Class II recall Z-0805-2013, published February 20, 2013 and initiated June 4, 2012. Recalling firm: Aesculap, Inc.; Status: Terminated; Reported quantity: 11; Lot: Lots 4503489878, 4503792749, 4503682927, 4503508385, 4503789879, 4503800248 and 4503682935. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in the ventricle. There is a small percentage of cases where aggressively attempting to adjust the deflector has a potential to cause the catheter
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Aesculap, Inc.
- Distributor
- Nationwide Distribution including MI, PA and TX.
- Lot
- Lots 4503489878, 4503792749, 4503682927, 4503508385, 4503789879, 4503800248 and 4503682935.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAColumbus Revision Knee System, EnduRo Knee System Product Usage: The Columbus Revision Knee System and EnduRoMarch 9, 2016
- FDAFlexible Screw Driver SJ706R; Non Sterile; B Braun; Aesculap; Aesculap AG These instruments are intended forDecember 30, 2015
- FDAS4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with aAugust 12, 2015
- FDAMiethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinalFebruary 4, 2015
- FDABREG Sterile Polar Pads; 02356 Rev M, PAD M/U XL STER, Mfg.Date: 112012. Model #s: 02510, 09901, 02330, 02350,February 20, 2013
- FDARemel Xpect Clostridium difficile Toxin A/B, IVD, 20 clostridium difficile toxin A/B tests, REF 24650. ProductFebruary 20, 2013
Questions about this recall
Was Aesculap(R) - Miethke Shunt System, FT040T - Control Reservoir Set w/Distal Catheter and FT075P 0 Ventric Cathet, w/Stylet and RT-ANG Guide Central Nervous System Shunt and Components intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. recalled?
Yes. FDA published recall Z-0805-2013 on February 20, 2013. Reported hazard: The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in the ventricle. There is a small percentage of cases where aggressively attempting to adjust the deflector has a potential to cause the catheter
What should I do if I have this product?
Voluntary: Firm initiated · Letter