Medtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01

Published February 20, 2019 · Recall initiated: January 17, 2019

Summary

FDA records medical device Class I recall Z-0808-2019, published February 20, 2019 and initiated January 17, 2019. Recalling firm: Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF); Status: Ongoing; Reported quantity: 7469 units; Lot: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01, GTIN: 00643169709003, All Serial Numbers (b) Model Number REDR01, GTIN: 00643169708990, 00643169969742, All Serial Numbers (c) Model Number REVDD01, GTIN: 00643169709010, 00643169969773, All Serial Numbers. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Distributor
Wordlwide
Lot
RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01, GTIN: 00643169709003, All Serial Numbers (b) Model Number REDR01, GTIN: 00643169708990, 00643169969742, All Serial Numbers (c) Model Number REVDD01, GTIN: 00643169709010, 00643169969773, All Serial Numbers
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01 recalled?

Yes. FDA published recall Z-0808-2019 on February 20, 2019. Reported hazard: A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality

What should I do if I have this product?

Voluntary: Firm initiated · Letter