Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

Published January 15, 2025 · Recall initiated: November 25, 2024

Summary

FDA records medical device Class II recall Z-0809-2025, published January 15, 2025 and initiated November 25, 2024. Recalling firm: Spacelabs Healthcare, Inc.; Status: Ongoing; Reported quantity: 2265; Lot: Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Spacelabs Healthcare, Inc.
Distributor
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic
Lot
Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data. recalled?

Yes. FDA published recall Z-0809-2025 on January 15, 2025. Reported hazard: Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.

What should I do if I have this product?

Voluntary: Firm initiated · Letter