TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.

Published January 31, 2024 · Recall initiated: December 18, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
St. Jude Medical, Atrial Fibrillation Division, Inc.
Distributor
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Lot
Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618.
Risk classification
Class II