TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
Published January 31, 2024 · Recall initiated: December 18, 2023
Summary
FDA records medical device Class II recall Z-0814-2024, published January 31, 2024 and initiated December 18, 2023. Recalling firm: St. Jude Medical, Atrial Fibrillation Division, Inc.; Status: Ongoing; Reported quantity: 483 units; Lot: Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- St. Jude Medical, Atrial Fibrillation Division, Inc.
- Distributor
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
- Lot
- Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACefixime for Oral Suspension 100mg/5mL, 50 mL bottle, Rx Only, Manufactured for: Lupin Pharmaceuticals, Inc.,February 7, 2024
- FDABudesonide Extended-Release Tablets 9mg, 30-count bottle, Rx Only, Manufactured by: Actavis Laboratories FL,February 7, 2024
- FDAbkActiv Ultrasound System 2300, Model numbers 2300-56 and 2300-66.January 31, 2024
- FDADiaSorin LIAISON HSV-2 Type Specific IgG, REF 310620January 31, 2024
- FDASmartPath to dStream for 1.5T, Model No. 781260, 782112January 31, 2024
- FDAIntera 1.5T, Model No. 781195January 31, 2024
Questions about this recall
Was TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile. recalled?
Yes. FDA published recall Z-0814-2024 on January 31, 2024. Reported hazard: When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.
What should I do if I have this product?
Voluntary: Firm initiated · Letter