Back to search
FDAFDA · Food and drugs·Recall number Z-0831-2013

LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct severe soft tissue and bony defects.

Published February 27, 2013 · Recall initiated: January 4, 2013

Summary

FDA records medical device Class I recall Z-0831-2013, published February 27, 2013 and initiated January 4, 2013. Recalling firm: DePuy Orthopaedics, Inc.; Status: Terminated; Reported quantity: 162 total units (US) and 102 units total (OUS); Lot: Part number: 198720028 and Lot numbers: 123510, 156211, 212355, B34GG1000, B4TH51000, B5RAX1000, B5RAXA000, B5RAXB000, CC3KX1000, CC3KXA000, E4RF31000, E4RF3A000, E55HV1, and EN4KN1000. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
DePuy Orthopaedics, Inc.
Distributor
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, FL, GA, ID, IL, MA, MD, MI, MN, NC, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV and the countries of Holland, Australia, Austria, South Africa and Canada including Veteran Administration hospitals.
Lot
Part number: 198720028 and Lot numbers: 123510, 156211, 212355, B34GG1000, B4TH51000, B5RAX1000, B5RAXA000, B5RAXB000, CC3KX1000, CC3KXA000, E4RF31000, E4RF3A000, E55HV1, and EN4KN1000.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct severe soft tissue and bony defects. recalled?

Yes. FDA published recall Z-0831-2013 on February 27, 2013. Reported hazard: DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients.

What should I do if I have this product?

Voluntary: Firm initiated · Letter