Back to search
FDAFDA · Food and drugs·Recall number Z-0831-2025

nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System

Published January 22, 2025 · Recall initiated: December 11, 2024

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Nuwellis Inc
Distributor
US Nationwide Distribution and the country of Hong Kong
Lot
Lot Numbers: 22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740.
Risk classification
Class I
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card