FDAFDA · Food and drugs·Recall number Z-0831-2025
nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System
Published January 22, 2025 · Recall initiated: December 11, 2024
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Nuwellis Inc
- Distributor
- US Nationwide Distribution and the country of Hong Kong
- Lot
- Lot Numbers: 22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States