nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System
Published January 22, 2025 · Recall initiated: December 11, 2024
Summary
FDA records medical device Class I recall Z-0831-2025, published January 22, 2025 and initiated December 11, 2024. Recalling firm: Nuwellis Inc; Status: Completed; Reported quantity: 845 units; Lot: Lot Numbers: 22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Nuwellis Inc
- Distributor
- US Nationwide Distribution and the country of Hong Kong
- Lot
- Lot Numbers: 22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAImpella RP Flex with SmartAssist; Product Number: 1000323;January 22, 2025
- FDABrand Name: Alcon Product Name: Monarch® III D Delivery Cartridge Model/Catalog Number: 8065977763 ProductJanuary 22, 2025
- FDAPROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), ELJanuary 22, 2025
- FDAAugustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1)January 22, 2025
- FDAMedline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527;January 22, 2025
- FDAX-Flow prostatectomy catheter, Device Reference Numbers AB6022, AB6024, AB6118, AB6120, AB6122, AB6124,January 22, 2025
Questions about this recall
Was nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System recalled?
Yes. FDA published recall Z-0831-2025 on January 22, 2025. Reported hazard: The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail