VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.
Published August 14, 2013 · Recall initiated: September 28, 2012
Summary
FDA records medical device Class II recall Z-0834-2013, published August 14, 2013 and initiated September 28, 2012. Recalling firm: Ventana Medical Systems Inc; Status: Terminated; Reported quantity: 54 instruments; Lot: Cat#750-601; PN 05987750001; Serial #'s: 310646, 311573, 310477, 311725, 311042, 311623, 312032, 312159, 310615, 310837, 311862, 311043 311572, 311766, 310526, 311011, 311185, 311723, 312082, 310635, 311574, 312084, 310636, 310835, 311179, 310493, 311183, 311769, 311930, 311962, 311963, 310437, 311965, 310997, 311726, 311620, 311512, 311526, 312083, 310468, 310759 (All serial numbers in the market). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ventana Medical Systems Inc
- Distributor
- Worldwide distribution: USA (nationwide) and country of: Germany.
- Lot
- Cat#750-601; PN 05987750001; Serial #'s: 310646, 311573, 310477, 311725, 311042, 311623, 312032, 312159, 310615, 310837, 311862, 311043 311572, 311766, 310526, 311011, 311185, 311723, 312082, 310635, 311574, 312084, 310636, 310835, 311179, 310493, 311183, 311769, 311930, 311962, 311963, 310437, 311965, 310997, 311726, 311620, 311512, 311526, 312083, 310468, 310759 (All serial numbers in the market).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVentana HE 600 System, automated slide preparer, for use in laboratories.July 20, 2022
- FDAVENTANA anti-Helicobacter Pylori (SP48) PAB 250, Catalog Number 06425623001, model 790-1015September 26, 2018
- FDAINFORM HPV III Fam 16 Probe - US Export, Catalog Number 05278856001, model 800-4295 Immunohistochemistry (IHC)September 26, 2018
- FDAiView DAB Detection Kit, Catalog Number 05266157001, model 760-091 Immunohistochemistry (IHC) for in vitroSeptember 26, 2018
- FDAVENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use Only Product Usage:August 14, 2013
- FDAAlaris PC unit, model 8015, with version 9.12-Keyboard Processor and logic board assembly part numberAugust 14, 2013
Questions about this recall
Was VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents. recalled?
Yes. FDA published recall Z-0834-2013 on August 14, 2013. Reported hazard: Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu
What should I do if I have this product?
Voluntary: Firm initiated · Letter