FDAFDA · Food and drugs·Recall number Z-0836-2013
Animas(R) 2020, Animas(R) IR 1250 (1200 plus), Animas(R) IR 1200 Insulin Infusion Pumps Indicated for continuous subcutaneous infusion of insulin.
Published February 27, 2013 · Recall initiated: December 17, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to an internal calendar date limit (December 31, 2015) these pumps will cease to operate on January 1, 2016, and will result in a call service alarm that is displayed on the pump display.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Animas Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) including Washington, DC, Puerto Rico, US Virgin Island and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Austria, Australia, Bermuda, Canada, New Zealand, Czech Republic, Italy, France, Finland, Hungary, Israel, United Kingdom, Spain, Sweden, Norway, Denmark, Germany, Mexico and Ireland.
- Lot
- All Animas 2020, IR1250 and IR1200 Pump Models - Part Numbers 101370-00, 101371-00, 101372-00, 101374-00, 101375-00, 100380-00, 100381-00, 100382-00, 100384-00, 100385-00, 100450-00, 100210-00, 100211-00, 100212-00, 100250-00, 100251-00, 100252-00, 100200-00, 100201-00, 100202-00, 100170-00, 100171-00 and 100172-00.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States