User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
Published February 5, 2014 · Recall initiated: October 7, 2013
Summary
FDA records medical device Class II recall Z-0851-2014, published February 5, 2014 and initiated October 7, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 2038; Lot: Material numbers 10284473, 10488224, 10444801, and 10444802. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use the incorrect Flex Cartridge to process the user defined method.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- USA (nationwide) Distribution including Puerto Rico.
- Lot
- Material numbers 10284473, 10488224, 10444801, and 10444802.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAKimberly-Clark RadiOpaque Radiofrequency Cannula, Gauge 22, Length 100 mm, Active Tip 10 mm, Distributed inFebruary 5, 2014
- FDAThe ACUTE Innovations Modular RibLoc System (RibLoc U plus) is comprised titanium plates and screws that areFebruary 5, 2014
Questions about this recall
Was User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. recalled?
Yes. FDA published recall Z-0851-2014 on February 5, 2014. Reported hazard: When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use the incorrect Flex Cartridge to process the user defined method.
What should I do if I have this product?
Voluntary: Firm initiated · Letter