Back to search
FDAFDA · Food and drugs·Recall number Z-0856-2015

Lite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is a software application developed by HORIBA Medical intended to receive, store, and manage Startup, Patient, and Quality Control (QC) results from an ABX Micros 60 Hematology Analyzer or ABX Micros CRP 200 Hematology Analyzer. Only one instrument connection is allowed per each Lite DM workstation. The Lite DM application runs on a PC within a Windows Operating System platform. Data transmitted from the instrument to the Lite DM workstation is a Serial RS-232 unidirectional ABX Format communication. The Lite DM can act as a standalone data manager for data received from the instrument, or allow transmission of this data to a Host (LIS  Laboratory Information System or EMR  Electronic Medical Record system). Following the completion of each blood sample analysis performed on the connected HORIBA Medical hematology analyzer, results are then communicated (uni-directional) to the Lite DM workstation. Results received will display to the user in a specific color coding format indicating any normal or panic reference range violations. Results are then rejected or accepted by the user for final reporting - formatted to guidelines stipulated by CLIA. At no time are any results interpreted; only reported for review and appropriate action.

Published December 31, 2014 · Recall initiated: November 19, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

HORIBA Medical is informing all HORIBA Medical Lite DM customers using software version 2.0.0.19 and lower of a software bug related to the potential for results received being applied to the wrong order in the Lite DM Worklist. This bug has not been seen in software versions above V2.0.0.19 IMPACT: Potential for results received being applied to the wrong order in the Lite DM Worklist. The is

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Horiba Instruments Inc
Distributor
Nationwide within the U.S.
Lot
Lot/Serial #(s): N/A Software Version(s): Versions 2.0.0.19 and lower. Expiration Date (s): N/A
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card