Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Published February 7, 2024 · Recall initiated: December 8, 2023
Summary
FDA records medical device Class II recall Z-0861-2024, published February 7, 2024 and initiated December 8, 2023. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 108 systems (14 US, 1 Canada, 93 ROW); Lot: (1) System Code: 722079; (1) UDI/DI: 00884838085268; (1) Serial Number: 1005, 866, 1361, 952, 1241, 1169, 756, 703351, 1135, 1122, 824, 748, 630, 638, 703385, 812, 703346, 1303, 1313, 893, 768, 929, 703638, 1357, 703810, 934, 1202, 703461, 703706, 1172, 703959, 703845, 703705; (2) System Code: 722224; (2) UDI/DI: 00884838099258; (2) Serial Number: 539, 395, 278, 134, 730, 252, 370, 36, 572, 645, 272, 495, 781, 62, 1388, 352, 1655, 690, 555, 895, 2165, 861, 60, 20, 648, 2231, 1778, 1431, 778, 477, 917, 1718, 388, 2042, 150, 764, 2036, 1421, 157, 879, 212, 144, 1463, 347, 575, 2187, 355, 2576, 1332, 424, 387, 863, 211, 474, 712, 549, 2309, 799, 2283, 639, 1700, 2071, 537, 1589, 2208, 2057, 771, 772, 1216, 1667. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distributor
- Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
- Lot
- (1) System Code: 722079; (1) UDI/DI: 00884838085268; (1) Serial Number: 1005, 866, 1361, 952, 1241, 1169, 756, 703351, 1135, 1122, 824, 748, 630, 638, 703385, 812, 703346, 1303, 1313, 893, 768, 929, 703638, 1357, 703810, 934, 1202, 703461, 703706, 1172, 703959, 703845, 703705; (2) System Code: 722224; (2) UDI/DI: 00884838099258; (2) Serial Number: 539, 395, 278, 134, 730, 252, 370, 36, 572, 645, 272, 495, 781, 62, 1388, 352, 1655, 690, 555, 895, 2165, 861, 60, 20, 648, 2231, 1778, 1431, 778, 477, 917, 1718, 388, 2042, 150, 764, 2036, 1421, 157, 879, 212, 144, 1463, 347, 575, 2187, 355, 2576, 1332, 424, 387, 863, 211, 474, 712, 549, 2309, 799, 2283, 639, 1700, 2071, 537, 1589, 2208, 2057, 771, 772, 1216, 1667;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAPhilips Azurion 5 M12. Model Number: 722227.November 4, 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.November 4, 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.November 4, 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.November 4, 2026
- FDAAzurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)February 7, 2024
- FDAsmiths medical Medfusion Model 3500 Syringe pumpFebruary 7, 2024
Questions about this recall
Was Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12) recalled?
Yes. FDA published recall Z-0861-2024 on February 7, 2024. Reported hazard: Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
What should I do if I have this product?
Voluntary: Firm initiated · Letter