EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.
Published March 2, 2016 · Recall initiated: January 21, 2016
Summary
FDA records medical device Class II recall Z-0864-2016, published March 2, 2016 and initiated January 21, 2016. Recalling firm: EKOS Corporation; Status: Terminated; Reported quantity: 10 units; Lot: Lot Number 150910004. Expiration date is August 15, 2018. Serial numbers: 001, 002, 003, 004, 005, 006, 007, 008, 009, 010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- EKOS Corporation
- Distributor
- Distributed in the states of AZ, CA, MA, MI, OH, and VA.
- Lot
- Lot Number 150910004. Expiration date is August 15, 2018. Serial numbers: 001, 002, 003, 004, 005, 006, 007, 008, 009, 010
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Symbiq infusion pump is a device used in a health care facility to pump fluids into a patient in aMarch 2, 2016
- FDAFlexor Check-Flo Introducer, Catheter introducer, Catalog #: KCFW-6.0-35-70-RB-HFANL0-HC Intended forMarch 2, 2016
- FDASyngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process,March 2, 2016
- FDAStryker, Gamma3 K-Wire with Washer, 13x3.2x500 mm Product Usage: Intended for the temporary stabilization ofMarch 2, 2016
- FDAThe EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power perNovember 13, 2013
- FDAEkoSonic MACH4 Endovascular Device (Catalog #500-55106, 500-55112; 500-55118; 500-55124; 500-55130; 500-55140;July 25, 2012
Questions about this recall
Was EkoSonic Endovascular System, Catalog Number 500-56112. It consists of three main components, a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and removable MicroSonic Device (MSD), and a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device. recalled?
Yes. FDA published recall Z-0864-2016 on March 2, 2016. Reported hazard: The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.
What should I do if I have this product?
Voluntary: Firm initiated · Letter