MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Published April 13, 2022 · Recall initiated: January 10, 2022
Summary
FDA records medical device Class II recall Z-0869-2022, published April 13, 2022 and initiated January 10, 2022. Recalling firm: Shimadzu Medical Systems; Status: Ongoing; Reported quantity: 12 systems; Lot: Model: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
What to do
FDA Mandated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Shimadzu Medical Systems
- Distributor
- U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A
- Lot
- Model: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADigital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40June 10, 2026
- FDAMODEL: X-RAY TV SYSTEM SONIALVISION G4April 13, 2022
- FDAPanther Fusion GBS Assay Cartridges 96 Tests, Catalog Number: PRD-04484. in vitro diagnostic test for group bApril 13, 2022
- FDACoryell Memorial Hosp Total, Joint Pack - Surgical KitApril 13, 2022
- FDAShimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems withJanuary 4, 2017
- FDAMobile X-ray system MobileDaRt Evolution/FDR Go Software The device is a mobile X-ray system and allows aDecember 30, 2015
Questions about this recall
Was MODEL: X-RAY TV SYSSTEM SONIALVISION safire17 recalled?
Yes. FDA published recall Z-0869-2022 on April 13, 2022. Reported hazard: It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
What should I do if I have this product?
FDA Mandated · Letter