Trilogy® Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Published December 31, 2014 · Recall initiated: December 4, 2014
Summary
FDA records medical device Class II recall Z-0875-2015, published December 31, 2014 and initiated December 4, 2014. Recalling firm: Zimmer Manufacturing B.V.; Status: Terminated; Reported quantity: 68 units; Lot: Item No. 00-6200-054-20; Lot 62714305. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Manufacturing B.V.
- Distributor
- Worldwide Distribution: US (nationwide) and countries of: France, Germany, Iran, Romania, Serbia, Singapore, Taiwan, and United Kingdom.
- Lot
- Item No. 00-6200-054-20; Lot 62714305
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVersaFx Femoral Fixation System Hex head compression screw 4.7 mm dia. 2 inch (51 mm) length 47116202400May 25, 2016
- FDAPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 14mm Sterile. ForApril 13, 2016
- FDAProduct 25 consists of all product under product code: LPH and same usage: Item no: 234703104 DISTAL MEDIALMarch 23, 2016
- FDAProduct 48 consists of all product under product code: JWH and same usage: Item no: 597206526 NEXGEN ALL-POLYMarch 23, 2016
- FDAJ-TAN D PD Drops (bromphenirame maleate 1 mg and pseudoephedrine hcl 7.5 mg/ teaspoon), 1 fl oz (30 mL)January 7, 2015
- FDASolitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl SleeveDecember 31, 2014
Questions about this recall
Was Trilogy® Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. recalled?
Yes. FDA published recall Z-0875-2015 on December 31, 2014. Reported hazard: Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit