Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Published December 31, 2014 · Recall initiated: December 4, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Zimmer Manufacturing B.V.
- Distributor
- Worldwide Distribution: US (nationwide) and countries of: France, Germany, Iran, Romania, Serbia, Singapore, Taiwan, and United Kingdom.
- Lot
- Item No. 00-6200-054-20; Lot 62714305
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States